CarieScan PRO - Using the PRO
Perform any cross-infection precautions normally taken in your clinic.
To ensure the accuracy of the detection of dental caries, the following steps must be taken prior to any measurements:
- Remove any intra-oral removable appliances.
- Clean all tooth surfaces to remove any plaque: Occlusal and free smooth surfaces should ideally be cleaned with a tooth brush or a prophylaxis brush and water. If toothpaste or prophylaxis paste is used care must be taken to ensure it is all removed prior to making a measurement in order to avoid false readings.
- Isolate the tooth to be measured and those in the immediate vicinity using cotton-wool rolls and a saliva ejector.
- Use the Tufted Sensor for measurement of occlusal and accessible smooth surfaces only.
- Open the blister pack by peeling off the Tyvek lid.
- After taking suitable cross-infection precautions hold the CarieScan with the display facing you and push the Tufted Sensor onto the CarieScan PRO with the tuft facing away from you, locating it with a click.
- Care must be taken if any plastic sheathing is used for cross-infection control purposes. Ensure that it does not prevent electrical contact between the end of the CarieScan PRO and the contacts inside the sensors.
- The Occlusal icon
and Smooth surface icon
will both be shown. - Ensure the surface of the tooth to be measured is free from any plaque and debris.
- With the Soft Tissue Contact attached, push the angled plug of the soft tissue cable into the socket at the side of the CarieScan PRO.
- Gently place the curled end of the Soft Tissue Contact over the lower lip of the patient.
- Dry the site to be measured for 5 seconds with a 3-in-1 syringe to remove any pooled fluid.
- Press the Enter button
to start measuring. - The blue LED on the pyramid will flash indicating that the CarieScan PRO is in measurement mode.
- Place the sensor firmly onto the chosen occlusal or smooth surface and a measurement will automatically be taken. During the measurement 4 audible tones will be heard before data is displayed.
- On completion of the measurement the result is presented by a further sequence of (number) beeps and the colour pyramid classification will light displaying the caries detection level, as described in Displays in Detail.
- At this point the sensor tuft can be removed from the tooth surface and the result will remain visible on the displays until the next tooth measurement is detected.
- Record the last result as required and select another site to measure.
- Should the sensor become contaminated or waterlogged, you should replace the sensor with a new one before continuing.
- The Measurement mode (ReadyScan) will automatically restart after 3 seconds, and the CarieScan PRO will detect when the sensor is on a tooth.
- Take a new measurement by returning to step 9. (Further drying may not be necessary if repeating measurements on the same site.)
- Terminate the measurement process by pressing any button on the Keypad for about 1s.
- When measurements on the patient are complete:
- To remove the sensor from the CarieScan PRO, squeeze the sensor on the side grips to release the catch and remove from the device.
- All sensors must be disposed as sharp contaminated waste. All such waste must be placed in a suitable receptacle marked as biohazard, conforming to the required regulations.
- The CarieScan PRO and Soft Tissue Cable should be wiped down with alcohol-based wipes.
Note
Drying time is based on standard air pressure of 40psi.
|
Note: |