System Test
We recommend that a daily test of the system is performed.
  1. Turn the PRO ‘ON’ by pressing and briefly holding the ‘CANCEL’ key.
  2. With the display facing you connect the Cable Test Adaptor to the PRO neck with the pin facing away from you, locating it with a click.
  3. ‘SYSTEM TEST’ is shown on the display indicating the device has detected the adaptor.
  4. Connect the Lip hook connector cable to the PRO by pushing the plug fully into the socket on the side of the PRO.
  5. Connect the Lip Hook connector cable socket onto the pin of the adaptor.
  6. Press ‘ENTER’ to start the test.
  7. ‘Ok’ indicates a pass.
  8. ‘CFL’ may indicate the lip hook connector cable is damaged. In this case replace the cable or contact your supplier for advice.
  9. The PRO will return to the ‘SYSTEM TEST’ state until the Cable Test Adaptor is removed.

 

 